New TopShim run:

version = topshim 2.0.0
date = 2024-08-12
time = 13:29:46.438
command = topshim  plotall convcomp ordmax=5

FINE SHIMMING

Parameters:

probe = Z8400_0264
solvent = CDCl3
shim nucleus = 2H
nucleus optimised for = 19F
o1p (from lock) = 7.24 ppm

Results:

NOTE: No 19F B1-profile found. Optimizing shimming as if for NMR nucleus = 1H


1D Shimming:
Iteration  1: FWHM Before= 0.6 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  2: FWHM Before= 0.4 Hz, Target= 0.3 Hz, After= 0.4 Hz
Iteration  3: FWHM Before= 0.4 Hz, Target= 0.4 Hz
Shim succeeded

Lineshape quality factor (1D spectrum, spinning) = 0.538

shim changes: 
Z	-87
Z2	-189
Z3	+110
Z4	-536
Z5	-1303

duration = 1 min 15 sec
completed successfully
finished Mon Aug 12 13:30:59 2024


========================================================================